Type: |
2074 (battery) 2174 (loop) 4-20mA 2274 (line) |
Accuracy: |
±.25% of span, terminal point |
Case Size: |
3˝, 41⁄2˝ |
Case Material: |
3˝ stainless steel, 4-1⁄2" fiberglass reinforced thermoplastic or black epoxy coated aluminum |
Case Encl. Rating: |
Weatherproof, IP65 |
Wetted Materials: |
17-4 stainless steel (sensor), 316 stainless steel (socket) |
Socket Size: |
1⁄4 or 1⁄2 NPT, JIS, DIN, SAE, (1⁄2 NPT only with 4-1⁄2" case, others on application) |
Socket Location: |
Lower, 3, 9 and 12 o'clock |
Ranges: |
15 psi/Vac. thru 20,000 psi (see engineering units below for other units) |
Operating Temp.: |
14/140°F (10/60°C) |
Temp. Error: |
(Zero & Span) .04%/°F (<.02%°F Typical) |
Storage Temp.: |
–4/158° (–20°/70°F) |
DISPLAY |
Type: |
LCD |
Display Digits: |
Five (5) |
Character Height: |
3" case: .60", 4-1⁄2" case: .88" |
Backlight: |
Optional |
Bar Graph: |
Yes |
Battery Life: |
3˝ >1000 hrs., 41⁄2˝ >3600 hrs. |
Agency Approvals: |
CE, FM (Intrinsically Safe Class I, Div 1) CSA and CENELEC (pending) |
KEYPAD FUNCTIONS |
On/Off: |
Manually turns unit on and off (auto off options in menu) |
Zero/Clear: |
Zeros display or clears min. and max. values when displayed |
Min/Max (down) Arrow Key: |
Stores min & max values, arrow key allows for scrolling through menu items |
Menu Key: |
Provides access to menu options |
Backlight (up) Arrow Key: (backlight optional) |
Manually turns backlight on and off (auto off options in menu), arrow key allows for five menu options. (up) arrow key allows for scrolling through menu options |
Enter: |
Selects items in the menu |
MENU MODE |
Engineering Units: |
10 units of measurement are available; psi, In. H2O (with three temp. options: 20°C, 60°F, 4°C), Ft. H2O, mPa, mBar, kPa, kg/cm2, Bar, inHg and mmHg |
Configuration Mode: |
Allows for changes to default settings of gauge |
(Config): |
Including zero disable feature |
Bar Graph (Graph): |
Allows for adjustment of bar graph and 4-20 (optional feature) |
Auto Off (Off): |
Allows for changes to auto off of gauge, five options: Never, 2 min., 5 min., 15 min., 30 min. |
Update Rate (Update): |
Four options: 100 ms, 200 ms, 500 ms, 1 sec |
Dampening (Damp): |
Six options: None, average, 2, 4, 6, 8 times per 100ms |
Backlight: |
Five options: Never, 10 sec., 30 sec., 1 min., 5 min. |
Field Recalibration: |
Allows for recalibration of zero, midscale and span (password protected) |