Jump to Navigation

Yokogawa CA71 Handy Multifunction Calibrator

Handy Multifunction Calibrator
CA71

The CA71 lets you handle regular tests on TCs, RTDs and various other types of sensors and instruments, as well as operation checks when a problem has occurred. By itself it can source signals for input to equipment, and check output signal from equipment. With the CA71, you don't need to worry about synchronizing operations with other setting devices. (TC and RTD measurement functions are CA71 only.)

● Simultaneous signal source and measurement capacity
● Wide array of useful functions
● Easy operation
● Online communication function

Select a link under the "item" tab below to order.

Cat #Supplied AccessoriesList Price
CA71None$1,537.00
CA71/SP1CA71/SP1 includes Communications cable, data logging softare, 120V power supply, magnetic backing instrument case, source/measure cables, NIST traceable certification.$1,886.80
Signal generating unit response time Approximately 1 second (time between start of voltage change and when voltage enters accuracy range)
Signal generating unit voltage limiter Approximately 32 V
Signal generating unit current limiter Approximately 25 mA
Divided output (n/m) function Output = setting × (n/m) n = 0–m; m = 1–19; n ≤ m
Auto-step output function n value sent automatically when n/m function selection is selected (two options: approximately 2.5 seconds/step or approximately 5 seconds/step)
Sweep function Sweep time (two options: approximately 16 seconds or approximately 32 seconds)
Memory function 50 value sets (generated and measured values are stored as value sets with the same address (up to 50 value sets can be stored))
Measuring unit maximum input Voltage terminal: 300 V AC Current terminal: 120 mA DC
Current terminal input protection Fuses: 125 mA/250 V
Measuring unit ground voltage Maximum 300 V AC
Measurement display updating rate Approximately once per second
Serial interface Enabled when communication cable (RS232) is connected; sold separately as optional accessory (CA71 only)
Display Segmented LCD (approximately 76 mm × 48 mm)
Backlight LED backlight; auto-off after one minute (from when LIGHT key is turned on)
Power supply Four AA alkaline batteries, or special AC adapter (sold separately)
Battery life Measurement off, output 5 V DC/10 kΩ or greater: Approximately 40 hours
Simultaneous signal generation/measurement, output 5 V DC/10 kΩ or greater: Approximately 20 hours
Simultaneous signal generation/measurement, output 20 mA/5 V: Approximately 12 hours (using alkaline batteries, with backlight off)
Consumed power Approximately 7 VA (using 100 V AC adapter)
Auto-power-off function Approximately 10 minutes (auto-power-off can be disabled through a DIP switch setting)
Applicable standards IEC61010-1, IEC61010-2-31
EN61326-1
EN55011, Class B, Group 1
Insulation resistance Across input terminal and output terminal, 500 V DC, 50 MΩ or greater
Withstand voltage Across input terminal and output terminal, 3.7 kVAC, for one minute
Operating temperature and humidity ranges 0–50°C, 20–80% RH (no condensation)
Storage temperature and humidity ranges -20–50°C, 90% RH or less (no condensation)
External dimensions (WHD) Approximately 190 × 120 × 55 mm
Weight Approximately 730 g (including batteries)
Standard accessories All of the following are included:
Signal generating lead cables (one red, two black): 98020
Measurement lead cables (one red, one black): RD031
Carrying case: 93016
Terminal adapter for CA71: 99021
User's manual: IM CA71-E
Fuse: A1501EF (for current terminal input protection)
Four AA alkaline batteries: A1070EB × 4
Optional accessories
(sold separately)
AC adapter: A1020UP (100 V AC power supply)
AC adapter: A1022UP (120 V AC power supply)
AC adapter: B9108WB (220–240 V AC power supply)
RJ sensor: B9108WA (For reference junction compensation)
Accessory carrying case: B9108XA
Communication cable: 91017
Spare parts Signal generating lead cables (one red, two black): 98020
Measurement lead cables (one red, one black): RD031
Carrying case: 93016
Terminal adapter: 99021
Fuse: A1501EF (for current terminal input protection)

Signal source: DCV, DCA, Resistance, RTD, TC (10 kinds), Frequency, Pulse
Measurement: DCV, DCA, ACV, Resistance, RTD, TC (10 kinds), Frequency, Pulse
Options: CA71/SP1 includes Communications cable, data logging softare, 120V power supply, magnetic backing instrument case, source/measure cables, NIST traceable certification.



Main menu 2